Sign In | Join Free | My enlightcorp.com |
|
Brand Name : Kacise
Model Number : KPT1000-DP
Place of Origin : China (Mainland)
Certification : certificate of explosion-proof, CE
MOQ : 1pcs
Payment Terms : T/T, Western Union, MoneyGram
Supply Ability : 1000 Pieces per Week
Delivery Time : 5-8 work days
Packaging Details : each unit has individual box and all boxes are packed in standard packages or customers requests available
Working pressure (silicone oil) : 16MPa 25MPa 42MPa Three Levels
Static pressure effect : ±(0.05%URL+0.075%Span)/ 10MPa
storage temperature : -40~+85℃ With Display: -20~+65℃
Power supply effect : ±0.001%/10V(12~36)V DC, Negligible
Range code 0B : (0~0.6)KPa~6KPa
Measurement medium : Measures The Pressure Of Liquid, Gas Or Steam
Overpressure effect : ±0.1%*Span
Reference accuracy : ±0.05%
KPT1000-DP Monocrystalline Silicon Intelligent Differential
Pressure Transmitter
1.Introduction
The KPT1000-DP single crystal silicon differential pressure transmitter uses a single crystal silicon sensor chip made of advanced German MEMS technology, a double diaphragm structure, and exquisite production technology to achieve high precision and ultra-high stability of the product. The embedded intelligent signal processing module achieves the perfect combination of pressure and temperature compensation, providing reliable measurement accuracy and excellent long-term stability under a wide range of static pressure and temperature changes.
The KPT1000-DP single crystal silicon differential pressure transmitter can be used to measure the liquid level, density, and pressure of liquid, gas, or steam, and convert it into a 4-20mA DC (superimposed HART communication) current signal output;
Remote and local parameter setting function configuration is realized through Hart or RS485 communication and panel buttons.
2.Features
3.Applications
Model | KPT1000-DP high-precision differential pressure transmitter | |||||||||
Measurement medium | Measures the pressure of liquid, gas or steam | |||||||||
Measurement type | Differential pressure measurement, absolute pressure measurement | |||||||||
Output signal | 4~20mA+HART protocol digital signal RS485 serial port signal | |||||||||
Measurement range | Range code | 0B | 0C | 0E | 0F | 0G | ||||
Range | (0~0.6)KPa~6KPa | (0~2)KPa~40KPa | (0~2.5)KPa~250KPa | (0~10)KPa~1MPa | (0~30)KPa~3MPa | |||||
Recommended compression ratio | 10:1 | 20:1 | 100:1 | 100:1 | 100:1 | |||||
Reference accuracy | ±0.05% | If TD>10 (TD=maximum range/adjustment range), then: | ±(0.005*TD)% | If the range is 0B product, TD>6, then: | / | |||||
±0.075% | ±(0.0075*TD)% | ±(0.125*TD)% | ||||||||
±0.1% | ±(0.01*TD)% | ±(0.166*TD)% | ||||||||
Ambient temperature effect | The total impact at (-25~+65)℃ is: ±(0.15*TD+0.05)%*Span The total impact at (-40~-25)℃ and (+65~+85)℃ is: ±(0.2*TD+0.05)%*Span | |||||||||
Overpressure effect | ±0.1%*Span | |||||||||
Static pressure effect | ±(0.05%URL+0.075%Span)/10MPa | |||||||||
Long-term stability | ±0.15%URL/10 years | |||||||||
Power supply effect | ±0.001%/10V(12~36)VDC, negligible | |||||||||
Range limit | It can be adjusted arbitrarily within the upper and lower limits of the range. It is recommended to select a range code with the lowest possible range ratio to optimize performance characteristics. | |||||||||
Zero setting | Calibration range ≥ Minimum range | |||||||||
Installation position effect | The change of the installation position parallel to the diaphragm surface will not cause zero point influence. If the installation position is not parallel to the diaphragm surface, it can be eliminated by calibration. No influence on the range | |||||||||
Alarm current | Offline | 3.6mA | ||||||||
Online | 21mA | |||||||||
Insulation resistance | 100MΩ or more to ground (DC 500V) | |||||||||
Insulation strength | To ground AC 500V 50/60Hz, 1 minute | |||||||||
Response time | The damping constant of the amplifier component is 0.1S, and the sensor time constant is: (0.1~1.6)S, depending on the range and range ratio. The additional adjustable time constant is: (0.1~60)S. The impact on nonlinear output (such as square root function) depends on this function and can be calculated accordingly. | |||||||||
Ambient temperature | -40~+85℃ With display: -20~+65℃ | |||||||||
Storage temperature | -40~+125℃ | |||||||||
storage temperature | -40~+85℃ With display: -20~+65℃ | |||||||||
Working pressure (silicone oil) | 16MPa 25MPa 42MPa three levels | |||||||||
Static pressure limit | From 3.5KPa absolute pressure to rated pressure, the protection pressure can be greater than 1.5 times the rated pressure and applied to both sides of the transmitter at the same time | |||||||||
Isolation diaphragm | Stainless steel 316L,Hastelloy C-276,Tantalum,Gold plated | |||||||||
Process connection material | Stainless steel 304, 316 optional | |||||||||
Nuts and bolts | Stainless steel (A2-70), galvanized carbon steel (12.8) | |||||||||
Transmitter housing | Aluminum alloy + epoxy resin (surface spraying) Stainless steel | |||||||||
Sealing ring | Nitrile-Butadiene Rubber(NBR),Fluorinated Rubber(FKM),Polytetrafluoroethylene (PTFE) | |||||||||
Filling | Silicone oil, edible oil, FL oil | |||||||||
Process connection | Both ends of the process connection flange have NPT1/4 and UNF7/16 internal threads | |||||||||
Protection level | IP65, IP67(explosion-proof) | |||||||||
Weight | 2.6Kg (without mounting bracket, process connection accessories) |
Power supply and load conditions:
5.Wiring
① | + | Power supply and HART digital communication output terminals |
② | - | |
② | - | External indicator (ammeter) wiring terminals |
③ | + |
6.Dimensions
7.Support and Services
Our Precision Pressure Sensor product is backed by comprehensive technical support and services to ensure that you have the best experience possible. Our team of experts is available to answer any questions you may have about the product, from installation to operation. We also offer maintenance and repair services to keep your sensor functioning optimally. In addition, we provide calibration and customization services to meet your specific needs. Trust us to provide the support and services you need to make the most of your Precision Pressure Sensor.
8.Packing and Shipping
Product Packaging:
Shipping:
![]() |
0-0.6KPa-6KPa Range Code 0B Smart DP Transmitter with ±0.05% Reference Accuracy Images |